Home » Special Bed Chromatography » Expandet Bed Adsorption

Expandet Bed Adsorption

Expanded Bed Adsorption Chromatography

Expanded bed adsorption (EBA) is a preparative chromatographic technique which makes processing of viscous and particulate liquids possible. Prof. Arwind Lali, Resindion and LCC formed 2002 a collaborative network to build a EBA Chromatography prototype to separate very large proteins.

What is so unique with EBA?

The protein binding principles in EBA are the same as in classical column chromatography and the common ion-exchange, hydrophobic interaction and affinity chromatography ligands can be used. After the adsorption step is complete, the fluidized bed is washed to flush out any remaining particulates.

Expanded Bed Chromatography Process
Expanded Bed Chromatography Process

 

Elution of the adsorbed proteins was commonly performed with the eluent flow in the reverse direction; that is, as a conventional packed bed, in order to recover the adsorbed solutes in a smaller volume of eluent. However, a new generation of EBA columns has been developed, which maintain the bed in the expanded state during this phase, producing high-purity, high yields of e.g. MAbs [monoclonal antibodies] in even smaller volumes of eluent. Process duration at manufacturing scale has also been cut considerably (under 7 hours in some cases).

EBA may be considered to combine both the “Removal of Insolubles” and the “Isolation” steps of the 4-step downstream processing heuristic. The major limitations associated with EBA technology is biomass interactions and aggregations onto adsorbent during processing.[2]

Where classical column chromatography uses a solid phase made by a packed bed, EBA uses particles in a fluidized state, ideally expanded by a factor of 2. Expanded bed adsorption is, however, different from fluidised bed chromatography in essentially two ways: one, the EBA resin contains particles of varying size and density which results in a gradient of particle size when expanded; and two, when the bed is in its expanded state, local loops are formed. Particles such as whole cells or cell debris, which would clog a packed bed column, readily pass through a fluidized bed.[3] EBA can therefore be used on crude culture broths or slurries of broken cells, thereby bypassing initial clearing steps such as centrifugation and filtration, which is mandatory when packed beds are used. In older EBA column designs, the feed flow rate is kept low enough that the solid packing remains stratified and does not fluidize completely. Hence EBA can be modelled as frontal adsorption in a packed bed, rather than as a well-mixed, continuous-flow adsorber.

An expanded bed chromatographic adsorption (EBA) column for a biochemical separation process comprises a pressure equalization liquid distributor having a self-cleaning function below a porous blocking sieve plate at the bottom of the expanded bed, an upper part nozzle assembly having a backflush cleaning function at the top of the expanded bed, a better distribution of the feedstock liquor added into the expanded bed ensuring that the fluid passed through the expanded bed layer displays a state of piston flow. The expanded bed layer displays a state of piston flow. The expanded bed chromatographic separation column has advantages of increasing the separation efficiency of the expanded bed.
Expanded-bed adsorption (EBA) chromatography is a convenient and effective technique for the capture of proteins directly from unclarified crude sample. In EBA chromatography, the settled bed is first expanded by upward flow of equilibration buffer. The crude feed, a mixture of soluble proteins, contaminants, cells, and cell debris, is then passed upward through the expanded bed. Target proteins are captured on the adsorbent, while particulates and contaminants pass through. A change to elution buffer while maintaining upward flow results in desorption of the target protein in expanded-bed mode. Alternatively, if the flow is reversed, the adsorbed particles will quickly settle and the proteins can be desorbed by an elution buffer. The mode used for elution (expanded-bed versus settled-bed) depends on the characteristics of the feed. After elution, the adsorbent is cleaned with a predefined cleaning-in-place (CIP) solution, with cleaning followed by either column regeneration (for further use) or storage.